Key Technical Specifications
| Parameter | Specification |
|---|---|
| Product Type | In-situ Zirconia Oxygen Analyzer |
| Measurement Range | Configurable O2 percentages |
| Max Process Temp | High-temperature rated |
| Output Signals | Analog loop with HART protocol |
| Relay Outputs | Isolated relay contacts |
| Probe Length | Configurable insertion depth |
| Mounting | Industry standard flanges |
| Calibration | Integrated automatic calibration |
| Operating Environment | Ruggedized for harsh areas |
Product Introduction
You don’t buy an oxygen analyzer to look pretty on a rack; you buy it because your boiler is wasting fuel and your emissions are creeping up. I’ve seen too many plants rely on extractive systems that clog with soot or lag significantly. The ABB Endura AZ20 is a direct-insert zirconia unit. It goes right into the flue. No sample lines to freeze, no pumps to rebuild.Engineers choose this specific AZ20 series because it handles the dirty work. It handles extreme process temperatures found in heavy combustion applications. The integrated automatic calibration is the real selling point. I don’t have to haul a cylinder of test gas up a ladder constantly. The transmitter handles the diagnostics and keeps the analog signal to the DCS honest. Just a heads-up: this specific revision is picky about its grounding. If you don’t bond the chassis properly, you’ll get noise on the HART loop.
Quality SOP & Tech Pitfalls
The Lab Report (SOP)
Before this unit ships, it goes through my personal gauntlet.
- Visual & Counterfeit Check: I inspect the casting, the nameplate, and the connector pins. The aftermarket is flooded with “compatible” sensors that fail prematurely.
- Live Test on Test Rack: We power it up, verify the DC input, and check the analog loop response using a precision decade box.
- Insulation & Multimeter Checks: Using a Fluke multimeter, I verify the heater circuit resistance and check for any shorts to ground in the transmitter board.
- Firmware/Serial Logging: We log the exact serial number and ensure the internal electronics match the specific configuration string.
- Anti-static Sealing: The analyzer is purged, capped, and sealed in anti-static packaging to prevent ESD damage to the HART transceiver during shipping.
The Engineer’s Warning (Pitfalls)
Let me save you a headache. The configuration string isn’t random. It dictates the probe length, flange type, and thermocouple specs. I once had a client order a standard AZ20 to replace a specific length configuration because “they looked the same.” The new probe was too short. It didn’t reach the center of the flue, read ambient air, and the boiler control loop went haywire. Always match the full configuration string. Also, failing to use anti-seize compound on the mounting bolts will result in a seized probe that you’ll have to cut out with a torch during the next outage.
Installation & Configuration Guide
Phase 1: Pre-Installation
- ️ Isolate the process. Lock out and tag out the draft fan. Ensure the flue is depressurized and purged.
- ️ Wait for thermal dissipation. If the boiler was running, the flue walls are hot. Don’t burn your hands.
- Photograph everything. Take high-res photos of the old wiring, the DIP switches on the transmitter board, and the HART configuration menu.
Phase 2: Removal
- Disconnect the DC power and the analog loop. Cap the wires immediately to keep moisture out.
- Disconnect the test gas lines (if equipped with auto-cal).
- Loosen the mounting flange bolts evenly. Support the probe’s weight as it breaks free. Pull it straight out to avoid damaging the zirconia cell.
Phase 3: Installation
- Clean the mounting surface. Any soot or old gasket material on the flange will cause a false air leak, skewing your O2 readings high.
- Apply anti-seize. Use a high-temperature copper or nickel anti-seize on the mounting threads.
- Insert the new AZ20 probe. Ensure the cell is facing the correct direction (usually into the gas flow).
- Torque the flange bolts evenly in a star pattern. Reconnect the electrical and test gas lines exactly as photographed.
Phase 4: Power-On & Testing
- Restore DC power. Watch the HMI display. The heater will pull maximum current to bring the zirconia cell up to operating temperature.
- Wait for the “Stable” or “OK” LED. The warm-up cycle takes time.
- Perform a manual span calibration using a known test gas to verify the transmitter is reading correctly before handing it back to the DCS.
Compatible Replacement Models
| Compatibility Tier | Model Number | Notes |
|---|---|---|
| Drop-in Replacement | AZ20 (Identical Config) | 100% identical. Match the configuration string exactly for zero downtime. |
| ️ Software Compatible | AZ20 (Different Config) | May require rewiring or HART recalibration if probe length or flange type differs. |
| Hardware Mod Required | AZ25 / AZ30 / AZ40 | Different series for extreme temps, HazLoc, or extractive setups. Requires new hardware. |
Frequently Asked Questions (FAQ)
Q: Can I hot-swap the electrical connector while the boiler is running?
A: Absolutely not. The transmitter electronics are susceptible to voltage spikes. Hot-plugging a DC connector will likely fry the internal board. Always kill the power at the breaker before touching the connector.Q: What happens if I don’t have the exact configuration string?
A: It’s a gamble. The base AZ20 might physically bolt on, but the configuration string dictates the probe length, flange standard, and thermocouple type. If you substitute it, you risk the probe not reaching the measurement point, leaking, or giving erratic readings. Stick to the exact part number.Q: Do I need to buy a separate amplifier for this?
A: No. The AZ20 has an integrated transmitter. You just feed it DC power and pull the analog signal and HART data directly to your PLC or DCS.Q: My reading is drifting high. Is the sensor dead?
A: Not necessarily. A high O2 reading usually means a false air leak. Check your flange gasket and the test gas lines. If the cell is actually degraded, the internal diagnostics will usually flag a “Sensor Degraded” or “High Internal Resistance” warning on the HMI.Q: Can I use this in an explosion-proof (HazLoc) area?
A: No. The standard AZ20 is not rated for hazardous areas. If your flue is in a classified zone, you must use the AZ30 explosion-proof model. Using a standard AZ20 in a HazLoc area is a major safety violation.